Description
Function |
Two, three, four-wire measure grounding resistance, soil resistivity; Ground voltage, AC voltage |
Ambient temperature and humidity |
23℃±5℃,below 75%rh |
Power Supply |
DC 9V 6-section LR14 Dry Battery Standby for more than 100 hours |
Interference voltage |
<20V(should be avoided) |
Interference current |
<2A (should be avoided) |
Measure R electrode spacing |
a>5d |
Measured electrode spacing |
a>20h |
Auxiliary Earth Resistance |
Reference Conditions <100Ω, Operating Conditions <5kΩ |
Range |
Grounding resistance: 0.00Ω to 30.00kΩ |
Soil resistivity |
: 0.00Ωm~9999kΩm |
Ground voltage |
: 0.00V to 600.0V |
Measurement Method |
Precision 4-wire, 3-wire measurement, simple 2-wire grounding resistance |
Measuring Methods |
Grounding resistance: rated current change method Soil resistivity: four-pole method Ground voltage: average rectification (S-ES interface) |
Test Frequency |
128Hz |
Short-circuit Test Current |
AC >20mA (sine wave) |
Open Circuit Test Voltage |
AC 28V max |
Electrode Spacing Range |
can be set from 1m to 100m |
Change Gear |
Grounding resistance: fully automatic shifting from 0.00Ω to 30.00kΩ |
Soil resistivity |
: 0.00 Ωm~9000kΩm automatic shifting |
Backlight |
Gray screen backlight, suitable for dim places |
Display Mode |
4-digital super-large LCD display, blue screen backlight |
Measuring Indicator |
During measurement, LED flash |
LCD Frame Dimension |
111mm×65mm |
LCD display area |
108mm×65mm |
Dimension |
L×W×H: 240mm×188mm×85mm |
Standard Test Wire |
4 wires: each for red 20m, black 20m, yellow 10m, and green 10m |
Simple Test Wire |
2 wires: each for yellow 1.5m and green 1.5m |
Auxiliary Earthing Rod |
4 PCS |
Measuring Rate |
Voltage to ground: about 3 times/second |
Earth resistance, soil resistivity |
: about 7 seconds/time |
Circuit Voltage |
below AC 600V(The ground voltage measurement function cannot be used to measure commercial power) |
Data Storage |
500 sets, “MEM” icon storage indicator, flash display “FULL” icon to indicate storage is full |
Data Read |
Data read function: “MR” icon display |
Overflow Display |
Exceeding measuring range overflow function: “OL” icon display |
Alarm Function |
Alarm when the measured value exceeds the alarm setting value |
Battery Voltage |
Low battery voltage symbol display |
Automatic Shut-down |
“APO” instruction, automatically shut down after 15 minutes of power on |
Power Consumption |
Standby: About 40mA (backlight off) |
Turn on and backlight on |
: about 43mA |
Measurement |
: about 75mA (backlight off) |
Weight |
Tester:1230g (including battery) |
Testing Wires |
:1300g |
Auxiliary earthing rod |
:720g(4 PCS) |
Working Temperature & Humidity |
-10℃~40℃;below 80%rh |
Storage temperature & humidity |
-20℃~60℃;below 70%rh |
Overload Protection |
Measuring earth ground resistance: between each interfaces of H-E、S-ES, AC 280V/3 seconds |
Insulation Resistance |
Over 20MΩ(between circuit and enclosure it is 500V) |
Withstanding Voltage |
AC 3700V/rms (Between circuit and enclosure) |
Electromagnetic Features |
IEC61326(EMC) |
Protection Type |
IEC61010-1 (CAT III 300V, CAT IV 150V, Pollution Degree 2); IEC61010-031; IEC61557-1 (grounding resistance); IEC61557-5 (soil resistivity); JJG 366-2004 |