In Schneider, we cover complete range of Low Voltage Switchgear including ACB’s, MCCB’s, MCB’s, MPCBs, Magnetic Contactors, Overload Relays, VFD’s, Soft Starters and Automation Devices etc.
Weidmuller provides Terminal Blocks, PCB Connectors, Marking Systems, Fieldbus Distribution Boxes, Solid State Relays, Power Supplies SPD’s, Enclosures and Cable Glands etc.
Hioki is a leading Electrical Measuring Instruments Manufacturer in the world providing AC/DC Clamp & Multimeters, Insulation & Earth Testers, Infrared Thermometers, Phase Detectors, Temperature & Humidity Loggers, Impedance Analyzers, Battery Testers, Power Meters, and Power Quality Analyzers etc.
Besides marketing genuine components we also provide customer support and after sales services. These two attributes are practiced by us in letter & spirit. This job is handled by a team of qualified engineers
Description
Measurement line type | Single-phase 2-wire, Single-phase 3-wire, Three-phase 3-wire or Three-phase 4-wire plus one extra input channel CH4 for voltage/current, (all channels AC/DC measurement) |
---|---|
Voltage ranges | Voltage measurement: 1000.0 V rms or DC, Transient measurement 2.200 kV peak |
Current ranges | 50.000 mA AC to 5.0000 kA AC, 10.000 A DC to 2.0000 kA DC (depends on current sensor in use) |
Power ranges | 50.000 W to 6.0000 MW (determined automatically based on current range in use) |
Basic accuracy | Voltage: ±0.2% of nominal voltage, Current: ±0.1 % rdg. ±0.1 % f.s. + current sensor accuracy, Active power: DC ±0.5 % rdg. ±0.5 % f.s. + current sensor accuracy, AC ±0.2 % rdg. ±0.1 % f.s. + current sensor accuracy |
Measurement items | 1. Transient over voltage : 200 kHz sampling 2. Frequency cycle : Calculated as one cycle 3. Voltage (1/2) RMS, Current (1/2) RMS: one cycle calculation refreshed every half cycle 4. Voltage swell, Voltage dips, Voltage interruption, RVC (Ver.2.0 or later) : Voltage (1/2) RMS calculation 5. Inrush current : half-cycle calculation: Calculated as the current RMS value for current waveform data sampled every half-cycle. 6. Frequency 200 ms: Calculated as 10 or 12 cycles 7. 10-sec frequency: Calculated as the whole-cycle time during the specified 10 s period 8. Voltage waveform peak, Current waveform peak 9. Voltage, Current, Active power, Apparent power, Reactive power, Active energy, Apparent energy, Reactive energy, Energy cost, Power factor, Displacement power factor, Voltage unbalance factor, Current unbalance factor 10. Voltage crest factor, Current crest factor 11. Harmonic/ Harmonic phase angle (voltage/ current), Harmonic power: 0 th to 50 th orders 12. Harmonic voltage-current phase angle: 1 th to 50 th orders 13. Total harmonic distortion factor (voltage/ current) 14. Inter harmonic (voltage/ current): 0.5 th to 49.5 th orders 15. K Factor (multiplication factor) 16. IEC Flicker, Δ V10 Flicker |
Record | Maximum recording interval: 1 year, Maximum number of recordable events: 9999 × 365 days |
Interfaces | SD/SDHC memory card, RS-232C (for communication/ LR8410 link), LAN (HTTP server/ FTP/ Send e-mail), USB 2.0 (for communication) |
Logger connectivity | Sends measured values wirelessly to logger by using a Bluetooth® wireless technology serial conversion adapter. (Supported devices: Hioki LR8410 Link-compatible loggers), Ver. 2.0 and later |
Display | 6.5-inch TFT color LCD (640 × 480 dots) |
Power supply | AC adapter Z1002 (100 V to 240 V AC, 50/60 Hz, rated current 1.7 A), Battery pack Z1003 (Continuous use: 8 hr, Charging time: Max. 5 hr 30 m with AC adapter) |
Dimensions and mass | 300 mm (11.81 in)W × 211 mm (8.31 in)H × 68 mm (2.68 in)D, 2.5 kg (88.2 oz) (including battery pack) |
Included accessories | Instruction manual ×1, Measurement guide ×1, Voltage cord L1000-05 ×1 set (Red/Yellow/Blue/Gray/Black, Alligator clip ×5, Spiral tube ×5), Color clip (for identifying clamp sensor color) ×1 set, Spiral tube ×5, AC adapter Z1002 ×1, Strap ×1, USB cable (1 m 3.28 ft length) ×1, Battery pack Z1003 ×1, PQ ONE (software, CD) ×1 |